No Current Sellers for Fluke 9150
Sign Up/Login to get notified when someone has one for sale.
Sign Up/Login to get notified when someone has one for sale.
Low-cost thermocouple furnace
NIST-traceable calibration included
RS-232 port standard
With a stability of ±0.5 °C, it has a temperature range to 1200 °C and a display accuracy of ±5 °C across its entire range.
With interchangeable temperature blocks, you can check thermocouples as small as 1/16 of an inch in diameter. The 9150 works with 115 or 230 VAC power.
The 9150 Thermocouple Furnace uses Fluke Calibration's own microprocessor-based controller for great stability and set-point accuracy. It has a removable well insert for versatility. It has rapid cool-down and heat-up times. And it comes with an RS-232 port for connection to a PC.
With a stability of 0.5 C, it has a temperature range to 1200 C and a display accuracy of 5 C across its entire range.
With interchangeable temperature blocks, you can check thermocouples as small as 1/16 of an inch in diameter. The 9150 works with 115 or 230 VAC power.
The 9150 Thermocouple Furnace uses Hart’s own microprocessor-based controller for great stability and set-point accuracy. It has a removable well insert for versatility. It has rapid cool-down and heat-up times. And it comes with an RS-232 port for connection to a PC.
You can now afford to check your thermocouples with this excellent cost-effective instrument. Why pay more for features you don’t need and can’t use? Each unit is factory-calibrated and comes with test data and a calibration traceable to NIST.
| General Specifications | |
| Temperature Range | 150 °C to 1200 °C (302 °F to 2192 °F) | 
| Display Resolution | 0.1 ° to 999.9 ° 1 ° above 1000 ° | 
| Stability | ±0.5 °C | 
| Display Accuracy | ±5 °C | 
| Well Diameter | 1.25 in (32 mm) | 
| Well Depth | 140 mm (5.5 in); (101 mm [4 in] removable insert plus 38 mm [1.5 in] insulator) | 
| Heating Time | 35 minutes to 1200 °C | 
| Cooling Time | 140 minutes with block | 
| Well-to-Well Uniformity | ±2.5 °C (Insert “C" at 1200 °C) | 
| Stabilization | 20 minutes | 
| Power | 115 V AC (±10 %), 10.5 A or 230 V AC (±10 %), 5.2 A, switchable, 50/60 Hz, 1200 W | 
| Size (HxWxD) | 315 mm x 208 mm x 315 mm (12.4 in x 8.2 in x 12.4 in) | 
| Weight | 13 kg (28 lb.) | 
| NIST-Traceable Calibration | Data at 150 °C, 300 °C, 450 °C, 600 °C, 800 °C, 1000 °C, and 1200 °C | 
There are no reviews for this product at the moment
    No current Alternatives.
    You may find Alternatives in category.