Yokogawa Insulation tester MY10-03

Product Details

  • Manufacturer: Yokogawa

  • Model: MY10-03

  • Description:

    Rating 500V/100M_; Effective Measuring Range 0.05-100M_; Central Scale Value 2M_; AC Voltage Measurement 0 to 500 V; Lower Limit 0.5M_; Rated Current 1-1.2 mA

No Current Sellers for Yokogawa MY10-03

Sign Up/Login to get notified when someone has one for sale.

Analog models with single ratingMY10-01:125V/20M_MY10-02:250V/50M_MY10-03:500V/100M_MY10-04:500V/1000M_MY10-05:1000V/2000M_
AC voltage measurement
Automatic discharge
A wide choice of accessories-Designed for shared use with the MY40

Specifications – Model MY10-01

Rating
125V/20M_

Effective Measuring Range
0.01-20M_

Central Scale Value
0.5M_

AC Voltage Measurement
0 to 250 V

Lower Limit
0.125M_

Rated Current
1-1.2 mA

Approx. Overall Scale Length
107 mm

Battery Life
Approx. 10 hours when continuously operated on manganese – oxide batteries with the pointer pointing to the central scale value.

Batteries
Four AA (R6P) batteries

Approx. Width
125 mm

Approx. Height
103 mm

Approx. Depth
53 mm

Approx. Weight
500 g

Compliance
EN61010-1:1993; EN61010-2-31:1995 (Overvoltage Category III, Pollution Degree 2 installations for indoor use)

Standard Accessories
Batteries (Qty – 4) Earth Probe (98002) (Qty – 1) Line Probe (98001) (Qty – 1) Protection Cover (93013) (Qty – 1) Shoulder Strap (99005) (Qty – 1) User’s Manual (Qty – 1)

Standard Test Conditions – Model MY10-01

Ambient Temperature/Humidity Ranges
23 ±5 /45-75% RH

Position of Use
Horizontal (5° max. of angle of inclination)

Battery Voltage
Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)

Tolerance Mentioned Conditions
Resistance measurement: First effective measuring range = ±5% of reading Second effective measuring range = ±10% of reading Infinite and zero scale values: 0.7% max. of scale length AC voltage: ±10% of maximum scale value No-load voltage: 130% max. of rated voltage Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range Short-circuit current: 12 mA max.

Additional Functions – Model MY10-01

Automatic discharge function

If the object under test remains electrified after the MEAS switch is turned off, the tester lets you know by turning on the LED lamp. If you leave the tester connected to the electrified object, the tester automatically begins to discharge electricity and then finishes discharging-the LED lamp comes on and then goes out.
When the object under test is capacitive and electrified, the tester lets you know by turning on the LED lamp. When left connected to the object, the tester automatically discharges electricity, thus preventing possible electric shock or spike noise at power-on.

Battery check (BAT mark on the scale plate)

Specifications – Model MY10-02

Rating
250V/50M_

Effective Measuring Range
0.01-50M_

Central Scale Value
1M_

AC Voltage Measurement
0 to 300 V

Lower Limit
0.25M_

Rated Current
1-1.2 mA

Approx. Overall Scale Length
107 mm

Battery Life
Approx. 10 hours when continuously operated on manganese – oxide batteries with the pointer pointing to the central scale value.

Batteries
Four AA (R6P) batteries

Approx. Width
125 mm

Approx. Height
103 mm

Approx. Depth
53 mm

Approx. Weight
500 g

Compliance
EN61010-1:1993; EN61010-2-31:1995 (Overvoltage Category III, Pollution Degree 2 installations for indoor use)

Standard Accessories
Batteries (Qty – 4) Earth Probe (98002) (Qty – 1) Line Probe (98001) (Qty – 1) Protection Cover (93013) (Qty – 1) Shoulder Strap (99005) (Qty – 1) User’s Manual (Qty – 1)

Standard Test Conditions – Model MY10-02

Ambient Temperature/Humidity Ranges
23 ±5 /45-75% RH

Position of Use
Horizontal (5° max. of angle of inclination)

Battery Voltage
Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)

Tolerance Mentioned Conditions
Resistance measurement: First effective measuring range = ±5% of reading Second effective measuring range = ±10% of reading Infinite and zero scale values: 0.7% max. of scale length AC voltage: ±10% of maximum scale value No-load voltage: 130% max. of rated voltage Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range Short-circuit current: 12 mA max.

Additional Functions – Model MY10-02

Automatic discharge function

If the object under test remains electrified after the MEAS switch is turned off, the tester lets you know by turning on the LED lamp. If you leave the tester connected to the electrified object, the tester automatically begins to discharge electricity and then finishes discharging-the LED lamp comes on and then goes out.
When the object under test is capacitive and electrified, the tester lets you know by turning on the LED lamp. When left connected to the object, the tester automatically discharges electricity, thus preventing possible electric shock or spike noise at power-on.

Battery check (BAT mark on the scale plate)

Specifications – Model MY10-03

Rating
500V/100M_

Effective Measuring Range
0.05-100M_

Central Scale Value
2M_

AC Voltage Measurement
0 to 500 V

Lower Limit
0.5M_

Rated Current
1-1.2 mA

Approx. Overall Scale Length
107 mm

Battery Life
Approx. 10 hours when continuously operated on manganese – oxide batteries with the pointer pointing to the central scale value.

Batteries
Four AA (R6P) batteries

Approx. Width
125 mm

Approx. Height
103 mm

Approx. Depth
53 mm

Approx. Weight
500 g

Compliance
EN61010-1:1993; EN61010-2-31:1995 (Overvoltage Category III, Pollution Degree 2 installations for indoor use)

Standard Accessories
Batteries (Qty – 4) Earth Probe (98002) (Qty – 1) Line Probe (98001) (Qty – 1) Protection Cover (93013) (Qty – 1) Shoulder Strap (99005) (Qty – 1) User’s Manual (Qty – 1)

Standard Test Conditions – Model MY10-03

Ambient Temperature/Humidity Ranges
23 ±5 /45-75% RH

Position of Use
Horizontal (5° max. of angle of inclination)

Battery Voltage
Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)

Tolerance Mentioned Conditions
Resistance measurement: First effective measuring range = ±5% of reading Second effective measuring range = ±10% of reading Infinite and zero scale values: 0.7% max. of scale length AC voltage: ±10% of maximum scale value No-load voltage: 130% max. of rated voltage Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range Short-circuit current: 12 mA max.

Additional Functions – Model MY10-03

Automatic discharge function

If the object under test remains electrified after the MEAS switch is turned off, the tester lets you know by turning on the LED lamp. If you leave the tester connected to the electrified object, the tester automatically begins to discharge electricity and then finishes discharging-the LED lamp comes on and then goes out.
When the object under test is capacitive and electrified, the tester lets you know by turning on the LED lamp. When left connected to the object, the tester automatically discharges electricity, thus preventing possible electric shock or spike noise at power-on.

Battery check (BAT mark on the scale plate)

Specifications – Model MY10-04

Rating
500V/1000M_

Effective Measuring Range
0.5-1000M_

Central Scale Value
20M_

AC Voltage Measurement
0 to 500 V

Lower Limit
1M_

Rated Current
0.5-0.6 mA

Approx. Overall Scale Length
107 mm

Battery Life
Approx. 10 hours when continuously operated on manganese – oxide batteries with the pointer pointing to the central scale value.

Batteries
Four AA (R6P) batteries

Approx. Width
125 mm

Approx. Height
103 mm

Approx. Depth
53 mm

Approx. Weight
500 g

Compliance
EN61010-1:1993; EN61010-2-31:1995 (Overvoltage Category III, Pollution Degree 2 installations for indoor use)

Standard Accessories
Batteries (Qty – 4) Earth Probe (98002) (Qty – 1) Line Probe (98001) (Qty – 1) Protection Cover (93013) (Qty – 1) Shoulder Strap (99005) (Qty – 1) User’s Manual (Qty – 1)

Standard Test Conditions – Model MY10-04

Ambient Temperature/Humidity Ranges
23 ±5 /45-75% RH

Position of Use
Horizontal (5° max. of angle of inclination)

Battery Voltage
Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)

Tolerance Mentioned Conditions
Resistance measurement: First effective measuring range = ±5% of reading Second effective measuring range = ±10% of reading Infinite and zero scale values: 0.7% max. of scale length AC voltage: ±10% of maximum scale value No-load voltage: 130% max. of rated voltage Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range Short-circuit current: 12 mA max.

Additional Functions – Model MY10-04

Automatic discharge function

If the object under test remains electrified after the MEAS switch is turned off, the tester lets you know by turning on the LED lamp. If you leave the tester connected to the electrified object, the tester automatically begins to discharge electricity and then finishes discharging-the LED lamp comes on and then goes out.
When the object under test is capacitive and electrified, the tester lets you know by turning on the LED lamp. When left connected to the object, the tester automatically discharges electricity, thus preventing possible electric shock or spike noise at power-on.

Battery check (BAT mark on the scale plate)

Specifications – Model MY10-05

Rating
1000V/2000M_

Effective Measuring Range
1-2000M_

Central Scale Value
50M_

AC Voltage Measurement
0 to 500 V

Lower Limit
2M_

Rated Current
0.5-0.6 mA

Approx. Overall Scale Length
107 mm

Battery Life
Approx. 10 hours when continuously operated on manganese – oxide batteries with the pointer pointing to the central scale value.

Batteries
Four AA (R6P) batteries

Approx. Width
125 mm

Approx. Height
103 mm

Approx. Depth
53 mm

Approx. Weight
500 g

Compliance
EN61010-1:1993; EN61010-2-31:1995 (Overvoltage Category III, Pollution Degree 2 installations for indoor use)

Standard Accessories
Batteries (Qty – 4) Earth Probe (98002) (Qty – 1) Line Probe (98001) (Qty – 1) Protection Cover (93013) (Qty – 1) Shoulder Strap (99005) (Qty – 1) User’s Manual (Qty – 1)

Standard Test Conditions – Model MY10-05

Ambient Temperature/Humidity Ranges
23 ±5 /45-75% RH

Position of Use
Horizontal (5° max. of angle of inclination)

Battery Voltage
Within effective voltage range (The pointer must stay within the range indicated by the BAT symbol when the battery check is performed.)

Tolerance Mentioned Conditions
Resistance measurement: First effective measuring range = ±5% of reading Second effective measuring range = ±10% of reading Infinite and zero scale values: 0.7% max. of scale length AC voltage: ±10% of maximum scale value No-load voltage: 130% max. of rated voltage Rated measuring current: 1 mA (0 to 20%) when in first effective measuring range Short-circuit current: 12 mA max.

Additional Functions – Model MY10-05

Automatic discharge function

If the object under test remains electrified after the MEAS switch is turned off, the tester lets you know by turning on the LED lamp. If you leave the tester connected to the electrified object, the tester automatically begins to discharge electricity and then finishes discharging-the LED lamp comes on and then goes out.
When the object under test is capacitive and electrified, the tester lets you know by turning on the LED lamp. When left connected to the object, the tester automatically discharges electricity, thus preventing possible electric shock or spike noise at power-on.

Battery check (BAT mark on the scale plate)

Product Average Rating:
(0 reviews)


There are no reviews for this product at the moment


Your Review: Please log in to add a review.

No current Alternatives.
You may find Alternatives in category.

Advertisement