Fluke ProSim 8 Vital Signs Simulator

Product Details

  • Manufacturer: Fluke

  • Model: ProSim 8

  • Description:

    The ProSim 8 tests ECG (including fetal ECG/IUP and arrhythmias), respiration, temperature, IBP/cardiac catheterization, cardiac output, NIBP and SpO2 

No Current Sellers for Fluke ProSim 8

Sign Up/Login to get notified when someone has one for sale.

ProSim 8 Vital Signs Patient Simulator

 

Designed to make comprehensive patient monitor testing fast and easy, ProSim 8 Vital Signs Simulator offers total patient monitor quality and safety testing from non-invasive blood pressure (NIBP) static pressure linearity and dynamic pressure repeatability tests, EC13 and ACLS ECG waveforms, to complete Nellcor and Nonin SpO2 low perfusion and Masimo Rainbow SET. Featuring specialized stay-connected ECG posts to ensure secure lead connections, ProSim 8 is the preferred tester for patient monitor quality assurance and safety professionals. 

The ProSim 8 tests ECG (including fetal ECG/IUP and arrhythmias), respiration, temperature, IBP/cardiac catheterization, cardiac output, NIBP and SpO2 (including Masimo multi-wavelength Rainbow SET)s in a single 5-minute PM testing tool. Wireless PC communication, customized presets, auto-sequences, barcode scanning, onboard memory, direct data capture, printing functions and single-step adjustability maximizes testing productivity; a customized carry case doubles as a mobile work station.

  • All-in-one patient simulator 80 % smaller and 17 lbs./7.7 kilos lighter than single function simulator test tools
  • 8-in-1 multifunction patient simulator is a combination ECG simulator, fetal simulator, arrhythmia simulator, respiration simulator, temperature simulator, IBP simulator, cardiac output simulator, cardiac catheterization simulator, NIBP simulator, SpO2 tester, and is the premier SpO2 simulator to test Rainbow multi-wavelength waveforms
  • Stay-connected ECG posts for easy/secure ECG snap and lead connections
  • Custom SpO2 r-curve for accurate testing of the latest and future oximetry technologies
  • Static pressure linearity testing
  • Repeatable NIBP simulation for dynamic pressure repeatability testing
  • Physiologically synchronized pulses across all parameters
  • Barcode scanning and direct data capture and printing functionality
  • Onboard, customizable patient pre-sets and auto sequences for fast/easy testing
  • Multi-language user interface offers choice of language selection
  • Integrated, easily-replaceable long-life battery
  • Optional PC-interface software offers customizable procedures/checklists and automated data capture/storage
  • Wireless communication for remote PC control of test device, as well as data transfer and automated regulatory reporting

 

 

 

General Specifications
Temperature
Operating: 10 °C to 40 °C (50 °F to 104 °F)
Storage: -20 °C to +60 °C (- 4 °F to 140 °F)
Humidity 10 % to 90 % non-condensing
Altitude 3,000 meters (9,843 ft)
Dimensions (L x W x H) 14.5 cm x 30.2 cm x 8.6 cm (5.7 in x 11.9 in x 3.4 in)
Display LCD color display
Communication
USB device upstream port Mini-B connector for control by a computer
USB host controller port Type A, 5 V output, 0.5 A max load. Connector for keyboard, barcode reader, and printer
Wireless IEEE 82.15.4 for control by a computer
Power Lithium-ion rechargeable battery
Battery Charger: 100 V to 240 V input, 15 V/2.0 A output. For best performance, the battery charger should be connected to a properly-grounded ac receptacle
Battery Life 9 hours (minimum), 100 NIBP cycles typical
Weight 1.87 kg (4.2 lb)
Safety Standards IEC/EN61010-1 3rd Edition; Pollution degree 2 CAT None
Certifications CE, CSA, C-TICK N10140 , RoHS
Electromagnetic Compatibility (EMC) IEC 61326-1:2006

 

Detailed Specifications
Normal-Sinus-Rhythm Waveform
ECG Reference The ECG amplitudes specified are for Lead II (calibration), from the baseline to the peak of the R wave. All other leads are proportional.
Normal Sinus Rhythm 12-lead configuration with independent outputs referenced to right leg (RL) Output to 10 Universal ECG Jacks, color-coded to AHA and IEC standards.
High-Level Output 0.2 V/mV ± 5 % of the ECG amplitude setting available on a BNC connector.
Amplitude 0.05 mV to 0.5 mV (0.05 mV steps) 0.5 mV to 5 mV (0.25 mV steps)
Other leads are proportional to Lead II (reference lead) in percentage per:
Lead I: 70
Lead II: 100
Lead III: 30
Lead V1: 24
Lead V2: 48
Lead V3: 100
Lead V4: 120
Lead V5: 112
Lead V6: 80
Amplitude Accuracy ± (2 % of setting + 0.05 mV)
ECG Rate 10 BPM to 360 BPM in 1 BPM steps
Rate Accuracy ±1 % of setting
ECG Waveform Selection Adult (80 ms) or neonatal (40 ms) QRS duration
ST-Segment Elevation Adult mode only. -0.8 mV to +0.8 mV (0.1 mV steps). Additional steps: + 0.05 mV and - 0.05 mV
Power-On Default 60 BPM, 1.0 mV, adult QRS and ST-segment elevation of 0 mV
Arrhythmia
Pacer Pulse
Amplitude 0 (off), ± 2, ± 4, ± 6, ± 8, ± 10, ± 12, ± 14, ± 16, ± 18, ± 20, ± 50, ± 100, ± 200, ± 500, and ± 700 mV for lead II (reference lead)
Accuracy Reference lead II: ± (5 % setting + 0.2 mV)
All other leads: ± (10 % setting + 0.4 mV)
Pacer Pulse Width 0.1 ms, 0.2 ms, 0.5 ms, 1 ms, and 2 ms ± 5 %
Paced Arrhythmias Atrial 80 BPM
Asynchronous 75 BPM
Demand with frequent sinus beats
Demand with occasional sinus beats
Atrio-ventricular sequential
Noncapture (one time)
Nonfunction
Power-On Default Amplitude 5 mV, width 1 ms, atrial waveform
Arrhythmia
Baseline NSR 80 BPM
PVC Focus Left focus, standard timing (except where specified)
Supraventricular Arrhythmia Atrial fibrillation (coarse or fine); atrial flutter; sinus arrhythmia; missed beat (one time); atrial tachycardia; paroxysmal atrial tachcardia; nodal rhythm; and supraventricular tachycardia
Premature Arrhythmia Premature atrial contraction (PAC); premature nodal contraction (PNC); PVC1 left ventricular; PVC1 left ventricular, early; PVC1 left ventricular, R on T; PVC2 right ventricular; PVC2 right ventricular, early; PVC2 right ventricular, R on T; and multifocal PVCs
Ventricular Arrhythmia PVCs 6, 12, or 24 per minute; frequent multifocal PVCs; bigeminy; trigeminy; multiple PVCs (one-time run of 2, 5, or 11 PVCs); mono-ventricular tachycardia (120 to 300 BPM in 5 BPM steps); poly-ventricular tachycardia (5 types); ventricular fibrillation (coarse or fine); and asystole
Conduction Defect First-, second-, or third-degree heart block; and right- or left-bundle-branch block
Advanced Cardiac Life Support
Shockable pulseless arrest rhythms Ventricular fibrillation (coarse), ventricular fibrillation (fine), unstable polymorphic ventricular tachycardia
Non-shockable pulseless arrest rhythms Asystole
Symptomatic bradycardia Sinus bradycardia (< 60 BPM)
2nd degree AV block, mobitz type I
2nd degree AV block, mobitz type II
Complete/3rd degree AV block
Right bundle branch block
Left bundle branch block
Symptomatic tachycardia: regular narrow-complex tachycardia (QRS < 0.12 seconds) Sinus tachycardia > 150 BPM
Supraventricular Tachycardia
Symptomatic tachycardia: regular wide-complex tachycardias (QRS > 0.12 seconds) Sinus tachycardia > 150 BPM
Supraventricular Tachycardia SVT with aberrancy
Irregular tachycardia Atrial fibrillation (coarse and fine), atrial flutter, unstable monomorphic ventricular tachycardia (120 BPM to 300 BPM), torsade de pointes/polymorphic ventricular tachycardia (long QT interval)
ECG Performance Testing
Amplitude (peak-to-peak) 0.05 mV to 0.5 mV (0.05 mV steps) 0.5 mV to 5 mV (0.25 mV steps) 
Other leads are proportional to Lead II (reference lead) in percentage per:
Lead I: 70
Lead II: 100
Lead III: 30
Lead V1 through V6: 100
Pulse Wave 30 BPM, 60 BPM, with 60 ms pulse width
Square Wave 0.125 Hz, 2 Hz, 2.5 Hz
Triangle Wave 0.125 Hz, 2 Hz, 2.5 Hz
Sine Wave 0.05 Hz, 0.5 Hz, 1, 2 Hz, 5 Hz, 10 Hz, 25 Hz, 30 Hz, 40 Hz, 50 Hz, 60 Hz, 100 Hz, and 150 Hz
R-Wave Detection
Waveform Triangular pulse
Rate 30 BPM, 60 BPM, 80 BPM, 120 BPM, 200 BPM, and 250 BPM
Width 8 ms to 20 ms in 2 ms steps, and 20 ms to 200 ms in 10 ms steps
Width accuracy ± (1 % of setting + 1 ms)
QRS Detection
Widths 8 ms to 20 ms in 2 ms steps, and 20 ms to 200 ms in 10 ms steps
Width accuracy ± (1 % of setting + 1 ms)
Rate 30 BPM, 60 BPM, 80 BPM, 120 BPM, 200 BPM, and 250 BPM
R-Wave up slope 0.875 amplitude, 0.4375 x width
R-Wave down slope Full amplitude, 0.5 x width
S-Wave up slope 0.125 amplitude, 0.0625 x width
Tall T-Wave Rejection
Waveform QT Interval 350 ms
T-Wave width 180 ms
T-Wave shape ½ sinewave
Amplitude 0 % to 150 % reference lead amplitude in 10 % steps
Rate 80 BPM
Rate Accuracy ± 1 % of setting
Amplitute Accuracy ± (2 % of setting + 0.05 mV)
ECG Artifact
Type 50 Hz, 60 Hz, muscular, baseline wander, respiration
Size 25 %, 50 %, 100 % of the normal sinus R-Wave for each lead
Lead Select All, RA, LL, LA, V1, V2, V3, V4, V5, V6
Fetal/Maternal ECG
Fetal Heart Rate (fixed) 60 BPM to 240 BPM in 1 BPM steps
Fetal Heart Rate (IUP) 140 BPM at beginning, then varies with pressure
Intrauterine-Pressure Waveforms Variable deceleration, early deceleration, late deceleration, and uniform acceleration
Wave Duration 90 seconds, bell-shaped pressure curve, from 0 mmHg to 90 mmHg and returning to 0
IUP Period 2 min, 3 min, or 5 minutes; and manual
Default Settings FHR 120 BPM, uniform deceleration wave, manual
Invasive Blood Pressure
Channels 2, each independently settable with identical parameters and are individually electrically isolated from all other signals
Input/Output Impedance 300 Ω - or ± 10 %
Exciter Input Range 2 to 16 V peak
Exciter-Input Frequency Range DC to 5000 Hz
Transducer Sensitivity 5 (default) or 40 µV/V/mmHg
Pressure Accuracy ± (1 % of setting + 1 mmHg) accuracy guaranteed for dc excitation only
Static Pressure - 10 to + 300 mmHg in 1 mmHg steps
Pressure Units mmHg or Kpa
Dynamic Waveforms
Types (default pressures Arterial (120/80)
Radial artery (120/80)
Left ventricle (120/00)
Right ventricle (25/00)
Pulmonary artery (25/10)
Pulmonary-artery wedge (10/2)
Right atrium (central venous or CVP) (15/10)
Pressure variability Systolic and diastolic pressures are independently variable in 1 mmHg steps
Swan-Ganz Sequence Right atrium, right ventrical (RV), pulmonary artery (PA), pulmonary artery wedge (PAW)
Cardiac Catheterization
Chambers Aortic, pulmonary valve, and mitral valve
Respiration Artifact
Arterial, radial artery, and left ventricle 5 % to 10 % multiplication
Other 5 mmHg or 10 mmHg
BP Output Circular DIN 5-Pin
Power-On Default 0 mmHg
Respiration
Rate 0 (OFF), 10 BrPM to 150 BrPM in 1 BrPM steps
Waves Normal or ventilated
Ratio (inspiration:expiration)
Normal 1:1, 1:2, 1:3, 1:4, 1:5
Ventilated 1:1
Impedance Variations (? Ω) 0.00 Ω to 1.00 Ω iin 0.05 Ω steps and 1Ω to 5 Ω in 0.25 Ω steps
Accuracy Delta ± (3 % of setting + 0.05 Ω)
Baseline 500 Ω, 1000 Ω (default), 1500 Ω, 2000 Ω, Leads I, II, III
Accuracy Baseline ±5 %
Respiration Lead LA or LL (default)
Apnea Selection 12 sec, 22 sec, or 32 seconds (one-time events), or continuous (Apnea ON = respiration OFF)
Power-On Default 20 BrPM, delta 1.0 Ω
Temperature
Temperature 30 °C to 42.0 °C in 0.5 °C steps
Accuracy ± 0.4 °C
Compatibility Yellow Springs, Inc. (YSI) Series 400 and 700
Output Circular DIN 4-Pin
Cardiac Output
Catheter Type Baxter Edwards, 93a-131-7f
Calibration Coeffecient 0.542 (0 °C injectate), 0.595 (24 °C injectate)
Blood Temperature 36 °C (98.6 °F) to 38 °C (100.4 °F) ± 2 % in 1 °C steps
Injectate Volume 10 cc
Injectate Temperature 0 °C or 24 °C
Cardiac Output 2.5, 5, 10 liters per minute ± 7.5 %
Faulty-Injectate Curve Waveform for simulation available
Left-To-Right-Shunt Curve Waveform for simulation available
Calibrated Pulse 1.5 ° for 1 second
Connector Circular DIN 7 pin
Power-On Default 5 liters per minute, 0 °C injectate, 37 °C blood temperature
Non-Invasive Blood Pressure
Pressure Units mmHg or kPa
Manometer (Pressure Meter)
Range 10 mmHg to 400 mmHg
Resolution 0.1 mmHg
Accuracy ± (0.5 % reading + 0.5 mmHg)
Pressure Source
Target pressure range 20 mmHg to 400 mmHg
Resolution 1 mmHg
NIBP Simulations
Pulse 2 mmHg max into 500 ml NIBP system
Volume of air moved 1.25 ml max
Simulations (systolic/diastolic [MAP]) Adult: 60/30 (40), 80/50 (60); 100/65 (77); 120/80 (93); 150/100 (117); and 200/150 (167) and 255/195 (215)
Neonatal: 35/15 (22); 60/30 (40); 80/50 (60); 100/65 (77); 120/80 (93) and 150/100
Pressure variability: systolic and diastolic pressures are variable by 1 mmHg
Repeatability Within ± 2 mmHg (at maximal pulse size independent of device under test)
Synchronization: normal Sinus heart rates: 30 BPM to 240 BPM Maximum rate at 1 ml: 240 BPM achievable with pulses up to 1 ml
Maximum rate at 1.25 ml: 180 BPM
Synchronization: Arrhythmias Premature atrial contraction (PAC), premature ventricular contraction (PVC), atrial fibrillation, and missed beat
Leak Test
Target pressure 20 to 400 mmHg
Elapse time 0:30 to 5:00 minutes: seconds in 30 second steps
Leakage rate 0 mmHg/minute to 200 mmHg/minute
Pressure Relief Test Range 100 to 400 mmHg
SpO2 Test (Optional)
% O2
Range 30 % to 100 %
Resolution 1 %
% O2 Accuracy
With oximeter manufacturer's R-curve Saturation within UUT specific range: ± (1 count + specified accuracy of the UUT)
Saturation outside UUT specific range: monotonic with unspecified accuracy
With Fluke Biomedical R-curves 91 % to 100 % ± (3 counts + specified accuracy of the UUT)
81 % to 90 % ± (5 counts + specified accuracy of the UUT)
71 % to 80 % ± (7 counts + specified accuracy of the UUT)
Below 7 % monotonic with unspecified accuracy
Heart Rate 30 BPM to 300 BPM in 1 BPM steps. SpO2 test is synchronized with ECG rate delayed by 150 ms, accuracy ±1% of setting
Transmission: ratio of detector current to LED current, expressed in parts per million (ppm)
Range 0 ppm to 300.00 ppm
Resolution 0.01 ppm
Accuracy + 50 %/- 30 % for compatible monitors, unspecified for others. Selected by finger size and color: dark, thick finger, medium finger, light, thin finger, neonatal foot. The full range and resolution are available in the engineering mode
Pulse Amplitude
Range 0 % to 20.00 %
Resolution 0.01 %
Artifact
Respiration Range: 0 % to 5 % of transmission
Resolution: 1 %
Rate: all ProSim respiration simulation settings
Ambient light Range: 0 to 5X transmitted light
Resolution: 1X
Frequency: DC, 50 Hz, 60 Hz, and 1 kHz to 10 kHz in 1 kHz steps
Masimo Rainbow Technology Test Masimo Rainbow technology with an optional adapter supplied by Masimo that allows the ProSim two wavelength to test the Rainbow multiple wavelength system
Compatible Manufacturer Products
With manufacturer R-curve Nellcor, Masimo, Nonin, and Nihon Kohden
With Fluke R-curve Mindray, GE-Ohmeda, Philips/HP, and BCI
 
 

Model Name Description
ProSim 8 
(3979409)
ProSim 8 Vital Signs Simulator US Includes:
  • ProSim 8 Getting Started Manual (3980667)
  • ProSim 8 User's Manual (3980671)
  • ProSim 8 Battery Pack (4021085)
  • USB Cable (4034393)
  • IBP Cable, unterminated (2392173)
  • ProSim 8 Carrying Case (4034597)
  • ProSim NIBP MANDREL SET (4308086)
  • AC/DC Power Supply
  • AC Power cord USA
  • Ansur demo CD-ROM
ProSim 8-01 
(3979411)
ProSim 8 Vital Signs Simulator Schuko Includes:
  • ProSim 8 Getting Started Manual (3980667)
  • ProSim 8 User's Manual (3980671)
  • ProSim 8 Battery Pack (4021085)
  • USB Cable (4034393)
  • IBP Cable, unterminated (2392173)
  • ProSim 8 Carrying Case (4034597)
  • ProSim NIBP MANDREL SET (4308086)
  • AC/DC Power Supply
  • AC Power cord Schuko
  • Ansure Demo CD-ROM
ProSim 8-02 
(3979427)
ProSim 8 Vital Signs Simulator UK Includes:
  • ProSim 8 Getting Started Manual (3980667)
  • ProSim 8 User's Manual (3980671)
  • ProSim 8 Battery Pack (4021085)
  • USB Cable (4034393)
  • IBP Cable, unterminated (2392173)
  • ProSim 8 Carrying Case (4034597)
  • ProSim NIBP MANDREL SET (4308086)
  • AC/DC Power Supply
  • AC Power cord UK
  • Ansur demo CD-ROM
ProSim 8-03 
(3979430)
ProSim 8 Vital Signs Simulator Japan Includes:
  • ProSim 8 Getting Started Manual (3980667)
  • ProSim 8 User's Manual (3980671)
  • ProSim 8 Battery Pack (4021085)
  • USB Cable (4034393)
  • IBP Cable, unterminated (2392173)
  • ProSim 8 Carrying Case (4034597)
  • ProSim NIBP MANDREL SET (4308086)
  • AC/DC Power Supply
  • AC Power cord Japan
  • Ansur demo CD-ROM
ProSim 8-04 
(3979448)
ProSim 8 Vital Signs Simulator Australia Includes:
  • ProSim 8 Getting Started Manual (3980667)
  • ProSim 8 User's Manual (3980671)
  • ProSim 8 Battery Pack (4021085)
  • USB Cable (4034393)
  • IBP Cable, unterminated (2392173)
  • ProSim 8 Carrying Case (4034597)
  • ProSim NIBP MANDREL SET (4308086)
  • AC/DC Power Supply
  • AC Power cord Australia
  • Ansur demo CD-ROM
ProSim 8-05
(3979453)
ProSim 8 Vital Signs Simulator Brazil Includes:
  • ProSim 8 Getting Started Manual (3980667)
  • ProSim 8 User's Manual (3980671)
  • ProSim 8 Battery Pack (4021085)
  • USB Cable (4034393)
  • IBP Cable, unterminated (2392173)
  • ProSim 8 Carrying Case (4034597)
  • ProSim NIBP MANDREL SET (4308086)
  • AC/DC Power Supply
  • AC Power cord Brazil
  • Ansur demo CD-ROM
ProSim SPOT
(3985658)
ProSim SpO2 Test Module
ProSim RAINBOW 
(4034609)
ProSim RAINBOW Test CABLE

 

Cables (IBP)
3010-0104 3010-0104 EM-1 BP cable (6ft) (6-pin) (part# 2198893)
For use with the following:
•GE Marquette Medical (PPG/E for M DR)
•Witt Biomedical
•Schiller PB Series
3010-0208 Nihon Kohden-NK-1 BP cable (5M) (5-pin) (part# 2462263)
(Requires Nihon Kohden Smart IBP Cable JP-910P for latest series TR-6000 and minDin to DIN adapter)
3010-0103 3010-0103 Datascope DS-1 BP cable (6F) (part# 2198887)
3010-0307 Fukuda Denshi (DS3300 series) FD-2 BP cable (12M) (12-pin orange) (part# 2199387)
3010-0357 GE Marquette Medical (Dash, Eagle, Solar, Tram and MacLab) MQ-3 BP cable (rectangular -11 M) (part# 2199627)
3010-0110 HP-3 BP cable (12M 5µV) 120 in, Mini-DIN to 12 pin HP/PH (part# 2198902)
For use with the following:
•Hewlett Packard/Philips (78-300, 78-500, 78-800, Merlin/Viridia/Omnicare)
(HP/Philips M1006B iBP module has a sensitivity of 5 uv/V/mmHg only. Select the HP-3 cable for this application)
3010-0111 HP-4 BP cable (12M 40µV) for ProSim 8, MedSim 300B, 217A (part# 2198916)
connecting to Hewlett Packard/Philips (78-300, 78-500, 78-800, Merlin/Viridia/Omnicare)
3010-0115 Mennen Medical (Horizon series) MM-1 (6M) (6-pin) (part# 2198933)
3010-0122 GE Marquette Medical (7000 and TRAM-AR series only) MQ-2 (8M round) (7-pin) (part# 2198978)
MiniDIN/DIN IBP Mini DIN to DIN IBP Adapter (part# 3408564)
•Nihon Kohden
Used to adapt PS420 or MPS450 blood pressure cables for use with ProSim series simulators.
3010-0116 PC-1 BP cable (6M) (6-pin) (part# 2198940)
For use with the following:
•Physio Control (VSM series)
•Medical Data Electronics (Escort series)
3010-0584 DX-1 BP cable (10F) (10-pin) (part# 2200955)
Used with the following:
•Puritan Bennett PB 240
•Datex (AS/3, CS/3, Compact, Cardio Cap II, Critical Care, Light)
3010-0248 QM-1 (6M) Blood pressure cable (10-pin orange) (part# 2198925)
for use with the following:
•Quinton (Q Cath series)
•Drager/Siemens
(Some of Drager/Siemens monitors use the standard 10 pin Siemen connector, but newer models require Drager's IBP Adapter 10 to 7 pin. Drager PN: 3368383)
3010-0366 SM-3 BP cable (15M) (part# 2199666)
•Siemens (Micor/Mingo)
3010-0114 SM-1 BP cable (10M) for Pro Sim 2/3, 4 or ProSim 8 (10-pin) (part# 2198925)
•Drager/Siemens (SIRECUST series)
(Use the SM-1 and the Siemens Medical Transducer Adapter (3368-383-E530U) to run a single invasive BP channel on the Siemens Medical SC6000 and SC9000 series monitors)
3010-0076 3010-0076 TK-1 (6M) BP Cable (part# 2198879) applies to: 
• BCI International 
• Criticare Systems Inc. (1100) 
• Critikon (Dinamap Plus) 
• Invivo Research 
• Ivy Biomedical (400 and 700 series) 
• North American Draqer (Vitalert 2000) 
• Protocol System (Propaq series) Zoll Defib 
• SpaceLabs (1050, 1700, PCMS series) 
(SpaceLabs adapters 700-0028-00 and 0120-0551-00 with TK1 used when testing the new UltraView Command Module) 
3010-0048FG Universal unterminated UU-1 BP cable (5-Pin DIN one end only) (part# 2392173)

 

ProSim 8 Cable Kits
3984993
PS8 DRAGER CABLES
Drager Infinity Cable Set (part# 3984993) for ProSim 8 includes:
•SM-1 BP Cable (part# 2198925, two) 
•UT-4,Low profile 1/4 inch phone plug 
•YSI 400 Series Compatible 2 conductor (part# 2523334)
3984968
PS8 GE CABLES
GE Marquette Eagle/Dash/Solar Cable Set (part# 3984968) for ProSim 8 includes:
•MQ-3 BP Cable (part# 2199627, two) 
•UT-4,Low profile 1/4 inch phone plug 
•YSI 400 Series Compatible 2 conductor (part# 2523334) 
•UT-2 Tamp Cable 700 Series YSI (part# 2199019) 
•PROSIM8-4402GECO Din cardiac Output Marq Eagle (part# 4022300)
3984922
PS8 HP/PHI CABLES
HP/Philips Intellivue Cable Set (part# 3984922) for ProSim 8 includes:
•HP-3 BP Cable (part# 2198902, two)
•HPT-2 Tamp/C.O. Injct Cable Assembly (part# 2199257)
•COA-1 Cable Assembly (part# 2199240)
•UT-4,Low profile 1/4 inch phone plug
•YSI 400 Series Compatible 2 conductor (part# 2523334)
3985009
PS8 NK CABLES
Nihon Kohden Cable Set (part# 3985009) for ProSim 8 includes:
•Nihon Kohden-NK-1, BP Cable (5M) (part# 2462263, two) 
•DIN to minDin adapter (part# 3408564) 
•UT-4, Low profile 1/4 inch phone plug 
•YSI 400 Series Compatible 2 conductor (part# 2523334)
3984910
PS8 ACC KIT
ProSim 8 Accessory Kit (part# 3984910) includes:
•DIN to minDin adapter (part# 3408564)
•HP/Philips Intellivue IBP cable (part# 2198902)
•GE Marquette Eagle/Dash/Solar IBP cable (part# 2199627) 
•Welch Allyn Propaq/SpaceLabs Ultraview IBP cable(part# 2198879) 
•USB wireless dongle (part# 3341333) 
•YSI400 series temperature cable (part# 2523334) 
•YSI700 series temperature cable (part# 2199019) 
•CI-3 Cardiac Output Box (part# 2392199) 
•spare battery pack (part# 4021085)
3984946
PS8 SPACEL CABLES
SpaceLabs Ultraview Cable Set (part# 3984946) for ProSim 8 includes:
•TK-1 BP Cable (part# 2198879, two) 
•UT-4,Low profile 1/4 inch phone plug 
•YSI 400 Series Compatible 2 conductor (part# 2523334) 
•UT-2 Tamp Cable 700 Series YSI (part# 2199019) 
•CI-3 Cable Assembly (part# 2392199)
3984979
PS8 WA/PRO CABLES
Welch Allyn/Propaq Cable Set (part# 3984979) for ProSim 8 includes:
•TK-1 BP Cable (part# 2198879, two) 
•UT-4,Low profile 1/4 inch phone plug 
•YSI 400 Series Compatible 2 conductor (part# 2523334) 
•UT-2 Tamp Cable 700 Series YSI (part# 2199019)

 

Batteries, Chargers, and Power Supply
BPPS68 ProSim 6/8 Battery Pack

 

Cables
4034393 Data Transfer Cable for use with ProSim 4, ProSim 8, ESA615
3010-0284 COA-1 Cardiac output adapter (Hewlett Packard)
1/4 Phone plug to 4 pin female audio connector
(HPT-2 also required for cardiac output simulation on HP patient-monitoring systems) (part# 2199240)
3010-0285 HPT-2 temperature adapter (2 pin) (part# 2199257)
(Hewlett Packard/Philips)(used with UT-1 for HP monitors)
(COA-1 also required for cardiac output simulation on HP patient-monitoring systems)
ProSim8-4402GECO ProSim8-4402GECO Cardiac Output Switch for GE (Marquette Eagle) (part# 4022300) for use with ProSim 2/3 and ProSim 8
3010-0193 UT-2 Standard 1/4" phone plug (compatible with YSI 700 series – 3 conductor) temperature cable (part# 2199019)
Use with the following:
•Spacelabs Ultraview (use adapter# 700-0031-00)
•GE Marquette (use adapter# 402015-005 or 402015-004)
•Welch Allyn/Propaq
UT-4 UT-4 Low profile, ¼ in phone plug, YSI 400 series compatible, two conductor (part# 2523334 or 3015280)
•HP/Philips 
•SpaceLabs Ultraview (adapter# 700-0031-00)
•GE Marquette Eagle/Dash/Solar (adapter# 402015-005 or 402015-004)
•Drager Infinity
•Welch Allyn/Propaq
•Nihon Kohden (requires Nihon Kohden Smart Temperature Cable K961 for the latest series TR-6000)

 

Miscellaneous Accessories
2719-0153FG 4-pin female audio connector (part# 2392158)
General purpose cardiac device output connector
3010-0289FG CI-3 Cardiac output test module (part# 2392199)
PROSIM 6/8 CASE ProSim CARRYING CASE (part# 4034597)
MANDREL-5001 ProSim NIBP Mandrel Set (part# 4308086)
ProSim Chamber NIBP Test 500ML Rigid Chamber (part# 4034611)